Metallography · hardness · mechanics · optics — one chainMetallography & testing
National High-TechShanghai SME2 Patents
7 products
HOW IT WORKS

How it works

εσReHRmFractureElastic regionGripspecimen

A universal testing machine drives the crosshead with a servo motor and records force–displacement from a load cell and extensometer, yielding yield strength ReH, tensile strength Rm, elongation after fracture A and Young’s modulus E. The elastic slope is stiffness, the yield plateau shows whether the material yields distinctly, and the post-necking drop corresponds to localised deformation. The ZHONGYAN WDW series, 0.5–300 kN dual-column, has a 1:100000 speed ratio and covers tension, compression, bending, shear and peel.

0.5–300 kN dual column1:100000 speed ratioTension, compression, bend
Generate a selection proposal
WORKFLOW

Standard workflow

The preparation/inspection workflow built around the Testing Machines — every step has its own consumables and parameters.

Workflow1Specimen prepMachine per GB/T …2GripWedge or hydrauli…3Set rateStress-rate contr…4LoadTo fracture with …5ReportAuto ReH, Rm, A, E
SPECIFICATIONS

Key specifications

Max force5 – 300 kN
Speed range0.005 – 500 mm/min
Displacement resolution0.001 mm
Force accuracy± 0.5 %
StandardsGB/T 228.1 · ISO 6892-1 · ASTM E8
SELECTION

How to choose

  1. 1

    Routine metal bar and plate acceptance → 100/300 kN dual column with wedge grips and a large-deformation extensometer.

  2. 2

    Plastics, rubber and film → low-force single-column machine with pneumatic grips to prevent slip; speed control accuracy matters most.

  3. 3

    High temperature, corrosion or fatigue → add an environmental chamber or furnace with dedicated fixtures, and check frame stiffness and travel margin first.

FAQ

FAQ

Fracture keeps occurring at the grip — is the data usable?
No — standards require fracture within the gauge length. Check grip slip, alignment and machining marks; reduce clamping force or switch to hydraulic grips.
No distinct yield point — the curve is continuous?
The material has no distinct yield (cold-worked steel, aluminium alloys). Report Rp0.2 proof strength instead and switch the software to the non-proportional-extension criterion.
Results scatter widely within one batch?
Check the specimens first (dimensional tolerance, roughness, inclusions), then the machine (calibration validity, consistent speed). Average at least three per batch and report standard deviation.