Metallography · hardness · mechanics · optics — one chainMetallography & testing
National High-TechShanghai SME2 Patents
HOW IT WORKS

How it works

θLight sourceReticleBeamsplitter物镜CCD / eyepiecePlane mirrorCollimated beam (round trip)

An autocollimator works on the principle that an object at the focal plane images back onto the focal plane: reticle lines are collimated by the objective into a parallel beam, reflect off a plane mirror, and are folded by a beamsplitter into the eyepiece or CCD, imaging back on the reticle. If the mirror tilts by θ, the image shifts by Δ = 2f·tanθ (f = focal length), so measuring image displacement gives angular displacement. The ZHONGYAN ZY series covers ±5″ to ±600″ with 0.1″ resolution for guideway straightness, table flatness, rotary-table indexing and machine-tool geometry.

Double-reticle autocollimation0.1″ resolution±5″ – ±600″ ranges
Generate a selection proposal
WORKFLOW

Standard workflow

The preparation/inspection workflow built around the Autocollimators — every step has its own consumables and parameters.

Workflow1Set upAlign instrument …2LevelRough-adjust unti…3Coarse readingRecord the initia…4Point-by-pointStep the mirror a…5EvaluateStraightness / fl…
SPECIFICATIONS

Key specifications

Measuring range± 5″ / ± 10″ / ± 60″ / ± 600″
Resolution0.1″ / 0.2″
Focal length200 – 500 mm
Repeatability≤ 0.3″
StandardsJJF 1110 · GB/T 20920
SELECTION

How to choose

  1. 1

    Straightness of guideways and granite tables → mid range (±60″) with a long focal length, stepping with a bridge plate.

  2. 2

    Rotary tables, index heads and optical angle calibration → short range with high resolution (±5″/0.1″) plus a polygon prism.

  3. 3

    Shop floor with long distances and large deviations → wide range (±600″) first, then re-measure on the high-precision range.

FAQ

FAQ

Readings drift and repeatability is poor?
First thermal equilibrium — soak the instrument and workpiece at the same temperature for 30 min. Then check mirror seating and floor vibration, and shade the reticle from direct light.
Over long distances the cross image blurs?
Usually air turbulence or an inadequate mirror flatness. Shorten the path and shield it from air flow; use a mirror flatter than λ/4, clean it and refocus.
Can it measure parallelism or squareness directly?
Yes, by differencing two measurements: measure the datum first for a reference angle, then the target face — the difference is parallelism (use a pentaprism for squareness).
Compare tray 0/4
Compare now